Introduction to Graphite Furnace
Atomic Absorption Spectrometry.
Introduction
You have already seen how atomic absorption can provide low level measurements
of elements using a flame atomization source. Replacing the flame with
a electrically heated carbon tube increases the concentration of atoms
in the light path which in turn allows much lower detection limits. Some
disadvantages of the technique are the long analysis times and the number
of variables that must be optimized. Due to time limitations, we will use
a cookbook method to determine lead concentrations in a variety of aqueous
samples.
Safety Precautions
-
Follow all instructions carefully
-
Obviously the furnace can be extremely hot. Do not touch the furnace or
surrounding area during operation.
-
Wear safety glasses at all times.
-
Take care to ensure that lamps are off before handling (hollow cathode
lamps run at high voltage)
The chemicals used in this experiment pose little hazard provided that
routine laboratory precautions are taken to avoid ingestion and skin contact.
Procedure
1. Preparation of Solutions
(Note: we are dealing with very low concentrations of a ‘sticky’ element,
therefore thorough cleaning and careful preparation of solutions is essential).
i. Calibration Standard. The use of an autosampler allows standards
to be made from a simple master standard. This certainly saves a lot of
time, however, if the master standard is incorrectly prepared all of the
results will be invalid. Prepare a master solution of 50 ppb lead from
the 1,000 ppm stock standard and deionised water.
ii. Modifier. Due to the volatile
nature of lead, a matrix modifier is required. The cookbook method calls
for 50 ?g of ammonium nitrate per firing. The autosampler will add 5?L
of matrix modifier per firing. Calculate the concentration of modifier
that you will need and prepare this solution in deionized water.
iii. Samples. All samples will be provided.
1. Instrument Setup and Optimization
i. Switch on Instrument and Gases. Open the argon gas cylinder
to 15psi (this should already be set so just open the main cylinder valve).
Switch on the main part of the spectrometer, using the switch at the back,
and the power supply for the graphite furnace (smaller box to the right).
Switch on the water chiller by pressing the i/o button. Ensure that DI
water is in the wash bottle and that the waste bottle is not full. Open
the Solaar software and click on the wizard button.
(Take a note of gas cylinder pressure at the beginning and end of your
experiment and notify the instructor if it is low)
-
Optimization. Open the ‘analysis wizard’ and follow the directions
given. Make sure the method ‘chem352Pb’ is loaded and hit ‘next’. Hit ‘optical
setup’ and ‘aurozero’ and proceed to the next page. You can assume that
the furnace position is already optimized. Click on the ‘GFTV’ button and
select ‘analysis’. This will open a program that allows you to view the
inside of the graphite cuvet. Now we can ‘align probe’ and view the probe
position in the GFTV screen. If the probe is correctly aligned it should
look like:
If the probe does not enter the cuvet you will need to adjust the autosampler
position using the control on the top and right side of the autosampler
unit until the probe reliably enters the cuvet. The depth of the probe
can be adjusted by a control on top of the probe itself. To clean the system
perform the following: ‘park autosampler’, ‘wash autosampler’, ‘purge autosampler’
and finally ‘clean cuvet’ which will heat the furnace to burn off any contaminants.
On the next screen select the ‘ASLG’ button and load the autosampler tray
as indicated.
2. Performing the analysis
Once the optimization is complete, you can start the analysis. The
instrument is totally automated so you don’t need to watch it, however,
keep an eye on the results periodically so that problems can be identified
and corrected as early as possible.
3. Shut Down and Print Out
Switch off the spectrometer, power supply and water chiller. Close
of the argon cylinder by the main valve. Your results can be printed by
selecting ‘options’ under the ‘results’ pull down menu and making sure
that your results only are in the right-hand column. Under the file menu
select ‘print options’ and check boxes for ‘large font’, ‘method’, ‘description’,
‘resample results’, and ‘calibration graph’only. You can now print
your results.
4. Report
In your report make sure to answer the following questions:
-
What is the purpose of a matrix modifier?
-
The manual for the instrument claims that
a 20?L sample of 1.7?g/L lead solution gives an absorbance of 0.1 Absorbance
Units. What is the concentration of lead that gives 0.1 Absorbance Units
using your calibration?